Vigilance Requirements: MDR Articles 87 - 90 ,IVDR Articles 82-85
Manufacturer will analyse and find out whether the
incidents are serious incidents or undesirable side-effects. Serious
incidents are needs to be reported to competent authority (CA) per MDR article 87
/ IVDR article 82. Non serious incidents or that are expected undesirable side-effects
shall be documented in trend reporting (MDR article 87/ IVDR article 83)
Timelines for reporting of serious incidents and field
safety corrective actions (MDR Article 87/IVDR Article 82)
Manufacturers needs to report serious incidents to
competent authorities once a causal relationship to the device has been
established.
·
Serious incidents (SI) -no later than
15 days
·
Serious public health threat -no
later than 2 days
·
Death or an unanticipated serious
deterioration in a person's state of health the report -no later than 10 days
·
Field safety corrective action (FSCA)
-Without undue delay
Manufacturer may submit an initial report that is
incomplete followed up by a complete report to CA. Final report is submitted by
manufacturer through electronic system. Final report shall have conclusions
about the incident investigation and corrective actions.
Periodic summary reports
Manufacturer shall submit this report when below
listed conditions are fulfilled
·
Similar serious incidents that occur
with the same device or device type and
·
These incident root cause has been
identified or a field safety corrective action implemented
· Incidents
are common and well documented
Manufacturer shall coordinate with CA and CA needs to
agree the Periodic summary report format, content and frequency .
Healthcare professionals, users and patients also
report the report suspected serious incidents to the CA.CA will inform the
manufacturer about this incident and manufacturer will analyse and takes
appropriate action if required
MDR Article 88/IVDR
Article 83 Trend
reporting
· Reported in electronic system
·
Frequency or severity of Not serious incidents or that
are expected undesirable side- effects
· These incidents have
significant impact on the benefit-risk analysis
· PMS plan shall have
effective methods and tools to establish any statistically significant increase
in the frequency or severity of incidents within time period
MDR Article 89/IVDR Article 84 Analysis of serious
incidents and field safety corrective actions
After reporting serious incident, the
manufacturer shall conduct risk assessment of the incident and take any
required field safety corrective action. This will be evaluated by CA.
The following factor are considered by CA while evaluating the SI & FSCA
·
Causality, detectability and
probability of recurrence of the problem
·
Frequency of use of the device
·
Probability of occurrence of direct
or indirect harm
·
Severity of that harm
·
Clinical benefit of the device
·
Intended and potential users, and population
affected
National competent authority or the EMA scientific
opinion is needed for Devices with incorporates, as an integral part, a
substance(MDR)/ companion
diagnostic(IVDR).
Field safety notice (FSN)
Manufacturer shall inform field safety corrective
action to users through FSN. FSN shall be written in official Union language or
languages determined by the Member State in which the field safety corrective
action is taken.
Before informing user, FSN shall be submitted to CA
.CA will provide the comments .
FSN shall be uploaded in the electronic system and it shall be
accessible to the public
FSN shall include the below information
·
Device identification including UDIs,
SRN
·
Reason for the field safety
corrective action
·
Malfunctioned device reference
·
Associated risks for patients, users
or other persons
·
Actions to be taken by users.
MDR Article 90/IVDR Article 85 Analysis of
vigilance data
Commission shall, in collaboration with the Member states
analyse the vigilance data to
identify trends, patterns or signals in the data that may reveal new
risks or safety concern.
if there any new risk arises then it will inform
manufacturer.
MDR Article 92/ IVDR Article 87 Electronic system
on vigilance and on post-market surveillance
The following reports are submitted electronically,
·
Serious incidents and field safety
corrective actions
· Periodic summary reports
· PSURs
· Trend Reports
·
Field safety notices
·
Information to be exchanged between
the competent authorities of the Member States and between them and the
Commission
References
·
REGULATION (EU) 2017/745 OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL
of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation
(EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council
Directives 90/385/EEC and 93/42/EEC
·
REGULATION (EU) 2017/746 OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL
of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive
98/79/EC and Commission Decision 2010/227/EU
No comments:
Post a Comment